Super Rapid Biological Indicator for H2O2 Airborne & Surface Disinfection, 1h Results, 50/box
Couldn't load pickup availability
Product code: BI-BioSurf
OEM: BT97
How it works
Each test set has two tubes. Tube A holds the spore carrier: open it inside the area to be disinfected and set the cap on the BioSurf Holder, carrier facing up, so the spores are exposed directly to the H2O2. Tube B holds the culture medium and stays closed in the holder during the cycle. After disinfection and aeration, the exposed cap goes onto Tube B and the indicator is read in a Bionova auto-reader incubator at 60 °C. A negative result after 1 hour confirms the H2O2 killed the spores.
Features
- 1-Hour Readout: fluorescence readout at 60 ± 2 °C in a Bionova auto-reader incubator, such as the MiniBio Auto-Reader.
- Spores: Geobacillus stearothermophilus (ATCC® 7953), 10^6 spores per carrier.
- Tests the Actual Process: bacteria on a surface are exposed directly to the hydrogen peroxide.
- Two-Tube Design: easy handling without cross-contamination.
- Built-in Chemical Indicator: the indicator on the Tube B label turns green to confirm exposure to hydrogen peroxide. This color change alone does not prove the process was successful.
- Optional Visual pH Confirmation: after 48 hours, the medium turns yellow if spores survived and stays purple if disinfection was successful.
- Readout Reliability: ≥ 97% reliability of the 1-hour readout compared with the 7-day visual readout.
- Placement Mapping: a diagram is included to mark where each indicator was placed in relation to the disinfection device.
- Shelf Life: 2 years from the date of manufacture under recommended storage conditions.
Required, sold separately:
Bionova MiniBio Auto-Reader Incubator
Holder for BioSurf Biological Indicator
Technical Specifications
| Process | Airborne and surface disinfection by vaporized and aerosolized Hydrogen Peroxide |
|---|---|
| Packaging | 50 (25 test sets: Tube A + Tube B) |
| Microorganism | Geobacillus stearothermophilus (ATCC® 7953) |
| Population | 10^6 Spores/carrier |
| Super Rapid Readout | 1 hour |
| Visual Confirmation | 48 hours |
| Compatible Incubators | Bionova® IC10/20FR, IC10/20FRLCD or MiniBio Auto-Reader incubators, at 60 ℃ |
| Shelf Life | Biological Indicators have an expiration date of 2 years from the date of manufacture when stored at recommended conditions. Do not use indicators after their expiration date. |
| OEM | BT97 |
Composition
- Bionova® BT97 BioSurf Biological Indicator consists of two tubes, A and B.
- Tube A contains a population of Geobacillus stearothermophilus ATCC® 7953 spores embedded in a stainless steel carrier, which is attached to the lid.
- Tube B contains a purple growth indicator medium within a glass ampoule and a chemical indicator of exposure in the external label.
Storage Note: Store in a dark place and at temperatures between 10-30 °C, 30-80 % relative humidity. Do not freeze. Do not store biological indicators near sterilizing agents or other chemical products.
Instructions for Use
- Inside the room to be disinfected open tube A and place the lid down and carrier up on the holder specially designed for the product. The rest of the tube may be discarded. Thus, Geobacillus stearothermophilus spores are exposed to evaluate the disinfection process. Place tube B without uncovering in the remaining position of the holder.
WARNING: Place the holder with the indicator in those areas that you consider a priori more inaccessible to the hydrogen peroxide. Places subject to minimal flows are suggested such as corners of the room, sectors in and around the equipment and spaces between disposable materials that will be used in the room.
To define the most difficult areas to decontaminate, it is recommended to use several indicators in the same process.
NOTE: A scheme of the room is attached to the product, where you can mark the position of the indicators with respect to the disinfection device. - Carry out the process of decontamination and aeration as usual.
- Start the Bionova® Auto-reader Incubator, select the temperature (60 °C) and the required readout program (1 hour).
IMPORTANT: Always wear both sterile clothing and personal protective equipment (glasses, mask, gloves) to enter the enclosed area. Avoid sudden movements of air. Do not reveal Bionova® BT97 BioSurf Biological Indicator until the incubator reaches the required temperature. - Verify that the chemical indicator printed on the label of tube B turned to green. A color change confirm that the indicator has been exposed to hydrogen peroxide.
IMPORTANT: This color change is not evidence that the process was successful. If the chemical indicator color has not changed, review the process. - Crush the ampoule contained in the biological indicator with an individual ampoule crusher or with the ampoule crusher placed within incubator´s incubation area. Then shake the tube down vigorously, until the medium reaches the base of the tube and soaks the filter entirely. Incomplete wetting of the filter may lead to an incorrect fluorescence readout.
- Remove the lid of tube B by taking the tube through the base. Discard it keeping the tube in vertical position.
- Place the exposed lid of tube A on tube B so that the carrier contacts the culture medium.
- Press the lid until the plastic tube is completely closed.
IMPORTANT: Use a non-processed biological indicator as a positive control every time a processed indicator is incubated. The positive control ensures that the incubation conditions were adequate. - Incubate the processed biological indicator along with the indicator used as a positive control for a maximum of 1 hour at 60 ± 2 °C for Super Rapid Readout.
Disposal
Discard biological indicators after use according to your country’s healthcare and safety regulations. The positive biological indicators can be autoclaved in a gravity air displacement steam sterilizer at 121 °C for 30 minutes, 132 °C for 15 minutes or 134 °C for 10 minutes; or in a dynamic air removal steam sterilizer at 132 °C for 4 minutes or 135 °C for 3 minutes.
Precautions
- Do not use Bionova® BT97 BioSurf Biological Indicators to control processes different from the indicated in this instructions for use.
- Do not reuse the biological indicators.
Scroll down for frequently asked questions
| Product Code | Description | Qty | Price | View |
| BI-BioSurf | Bionova BioSurf Super Rapid Biological Indicator for H2O2 Airborne & Surface Disinfection, 1-hour results | 50/box (25 test sets) | $329.00 | Add To Cart |