Once-A-Day Bowie & Dick Test Pack Chemical Indicator with warning sheet designed for testing air removal efficiency of vacuum-assisted steam sterilizers, 20/box
Bowie-Dick Test Packs
Product code: BD-CD-TP
OEM: BD125X/1
Ensure daily usage for accurate sterilization assessment, and be cautious not to use in unsuitable sterilization cycles.
Features
Bowie-Dick Test Packs
Product code: BD-CD-TP
OEM: BD125X/1
Ensure daily usage for accurate sterilization assessment, and be cautious not to use in unsuitable sterilization cycles.
Features
-
Targeted Sterilization Assessment:
- Specifically designed for evaluating air removal efficiency in dynamic-air-removal steam sterilizers at 132 °C for 4 minutes and 134 °C for 3.5 minutes
- Specifically designed for evaluating air removal efficiency in dynamic-air-removal steam sterilizers at 132 °C for 4 minutes and 134 °C for 3.5 minutes
-
Comprehensive Packaging:
- Consists of chemical indicators on two test sheets (125 x 113 mm) positioned between porous materials and a reticulated foam, forming a complete test package
- Consists of chemical indicators on two test sheets (125 x 113 mm) positioned between porous materials and a reticulated foam, forming a complete test package
-
Visual Indicator Transformation:
- Primary test sheet with a regular pattern transforms from yellow to a uniform dark brown/black color, signaling successful air removal. A lighter-colored area indicates air removal failure
- Primary test sheet with a regular pattern transforms from yellow to a uniform dark brown/black color, signaling successful air removal. A lighter-colored area indicates air removal failure
-
Early Warning System:
- Includes a warning sheet with an indicator circle, turning from yellow to dark brown/black for early detection of air removal failures before appearing on the primary test sheet
- Includes a warning sheet with an indicator circle, turning from yellow to dark brown/black for early detection of air removal failures before appearing on the primary test sheet
-
Standards Compliance:
- Bowie-Dick Test Pack complies with ANSI/AAMI/ISO 11140-1:2014 and ANSI/AAMI/ISO 11140-5:2007 standards for reliability and accuracy
- Bowie-Dick Test Pack complies with ANSI/AAMI/ISO 11140-1:2014 and ANSI/AAMI/ISO 11140-5:2007 standards for reliability and accuracy
-
Daily Use Recommendation:
- Perform Bowie-Dick tests daily, before the first load, after installation or relocation, after malfunctions, process failures, and major sterilizer repairs
- Perform Bowie-Dick tests daily, before the first load, after installation or relocation, after malfunctions, process failures, and major sterilizer repairs
- Critical Warning on Time Extension:
- Avoid extending the recommended test time to prevent false endpoints due to entrapped air diffusion. Not suitable for gravity displacement steam sterilization cycles.
The Chemdye Bowie-Dick Test Pack not only complies with standards but also offers equivalent performance to industry benchmarks.
Instructions for use
| Process | Steam |
|---|---|
| Packaging | 20 |
| Initial Color | Yellow |
| Final Color | Black, Dark Brown |
| Indicator Type | Type 2 Chemical Indicator |
| Regulations | ISO 11140-1, ISO 11140-5 |
Instructions for use
1. Run a warm-up cycle to properly heat the sterilizer.
2. Place the Bowie-Dick Test Pack horizontally with the label facing up in an empty vacuum assisted sterilizer chamber, in the bottom/front of the sterilizer rack and over the drain.
*The Test Pack should read “THIS SIDE UP” when placed correctly.
*The Test Pack should read “THIS SIDE UP” when placed correctly.
3. Run the sterilizer cycle for 3.5 min. – 134 ºC / 4 min. – 132 °C.
4. Remove the test pack from sterilizer and allow it to cool down. Check color change of type 1 indicator on the label.
5. Unwrapped the pack and remove the test and warning sheets. Check them for color change and record the results.