Holder for Splat Cleaning Indicator, 1
Regular price
$28.00
Sale price
Regular price
Shipping calculated at checkout.
Couldn't load pickup availability
Chemdye Holder for cleaning and thermodisinfection Indicators, Splat monitor test used in Washer Disinfector, 1/each
Note: Compatible with all Tuttnauer Cleaning indicators.
Product code: C1-CDS-WDH
Characteristics:
- Chrome plating SUS304 grade stainless steel re-usable holder of great resistance and durability.
- Base: 32 mm wide, 72.5 mm long, pin at the back side.
- Basket: 32 mm wide, 37 mm long, 0.5 mm thick.
The holder is designed to mimic the cleaning challenges posed by a large process load during the washing cycle and can be fixed on the tray, which allows reproducible location of the cleaning indicator in the washer-disinfector.
Sold separately : C1-CDS - Chemdye Splat cleaning Indicators performance test for monitoring automatic Washer Disinfectors and Ultrasonic cleaners
Quanity Discounts are applied in the cart.
When placed within the holder, half of the test soil is obscured, thus simulating the joint of a surgical instrument, with water stream hitting it indirectly.
The other half of the test soil is covered by a mesh, thus posing a difficulty for water stream to reach it, mimicking shadowed or occluded surfaces.
The new design of holder optimizes the control of the washerdisinfector`s ability to force water and detergent into and onto the test soil.
Instructions for use
- Choose the Chemdye Splat indicator according to the desired difficulty.
- Place the cleaning indicator in side the holder.
- Place the system on the tray along with the material.
- Once the cleaning cycle is over, remove the indicator.
- Analyze indicators using the reference table and record results.
If unfavorable results are obtained, perform the appropriate corrective actions and repeat the washing process.
Refer to chart