Rapid Self-Contained Biological Indicator for Steam, 3h Results, 50/box
BT220
Regular price
$192.00
Sale price
Regular price
Shipping calculated at checkout.
Couldn't load pickup availability
Bionova Rapid Self-Contained Biological Indicator for Steam sterilization processes with completed results in 3 hour, 50/box
Biological Indicator Inoculated with viable Geobacillus stearothermophilus bacterial spores.
FDA, Rapid Fluorescence
Product code: BI-BS-3
Mfr: BT220
Features
Sold Separately:
Bionova Incubator
Bionova MiniBio Auto-Reader Incubator
Biological Indicator Inoculated with viable Geobacillus stearothermophilus bacterial spores.
FDA, Rapid Fluorescence
Product code: BI-BS-3
Mfr: BT220
Features
- 3 hour readout time for steam sterilization process.
- Geobacillus Stearothermophilus (ATCC® 7953) spores with a concentration of 10^6 spores per vial.
- Optional Visual pH Confirmation:
- Users have the option to perform a visual pH color change confirmation after 48 hours, enhancing flexibility and offering an additional layer of confirmation.
-
Cultural Medium Color Indication:
- The culture medium undergoes a color change, initially to green and then yellow, if the sterilization process is unsuccessful.
- Conversely, a successful process maintains the medium in a purple hue, enabling clear differentiation.
-
Dual D-Value Information:
- The product provides D-values for two temperatures (121 ºC and informed values at 132ºC and 135ºC), ensuring comprehensive data for different sterilization conditions.
-
Fluorescence Detection Capability:
- The inclusion of fluorescence detection allows for an alternative and rapid assessment of sterilization effectiveness, adding an extra layer of reliability.
-
Humidity Consideration for 7-Day Readout:
- When opting for the 7-day readout, users are advised to create a humidified environment to prevent the medium from drying out, ensuring accurate and reliable results.
-
Shelf Life:
- 2 year expiration date from the date of manufacture under recommended storage conditions.
-
User-Friendly Instructions:
- The product comes with clear and concise instructions for use, including labeling requirements, packaging guidelines, and post-sterilization procedures.
Sold Separately:
Bionova Incubator
Bionova MiniBio Auto-Reader Incubator
- 50 Biological Indicators per box
Storage Guidelines
- Store in a dark place within the temperature range of 10-30 °C and 30-80% relative humidity.
- Avoid freezing and keep biological indicators away from sterilizing agents or other chemical products.
Instructions for Use:
-
Labeling SCBI
- Identify SCBI with sterilizer number, load number, and processing date.
-
Packaging and Placement
- Pack SCBI with materials for sterilization, placing it in less exposed areas.
-
Sterilization Process
- Carry out standard sterilization procedures.
-
Post-Sterilization Steps
- After sterilization, open the door, wait five minutes, and carefully remove SCBI.
- Caution: Wear safety glasses and gloves.
- Warning: Avoid excessive handling to prevent glass ampoule bursting. Let SCBI cool to room temperature.
-
Process Indicator Check
- Inspect the Process Indicator on SCBI label. Brown color indicates exposure to steam.
- Note: Color change does not confirm sterility; further checks needed if no change.
-
Ampoule Crushing and Incubation
- Seal the tube's lid and crush the ampoule using an ampoule crusher.
- Shake the tube to soak the spore carrier completely.
- Place SCBI in the incubator.
- Important: Use a non-sterilized SCBI as a positive control daily. Ensure both controls are from the same batch.
-
Incubation and Readout
- Incubate processed and positive control indicators for a maximum of 3 hour at (60 ± 2) °C.
- Note: Incubation should not exceed a 7-day period after sterilization.
- Failure in sterilization is indicated by fluorescence detection.
- Negative result if no fluorescence at 3 hour
- Positive control must show positive fluorescence readout.
- Record positive results and discard SCBI promptly as instructed.
Precautions:
WARNING: Do not use SCBI to control sterilization processes other than
those specified.
WARNING: Place one or more SCBI in hard-to-reach areas to ensure complete
chamber sterilization. Evaluate load configurations to identify and place
SCBI in all such locations.
WARNING: Do not reuse SCBI.
WARNING: Do not reuse the sterilizer until the SCBI test result is
negative.