Skip to content

Free Ground Shipping on Orders Over $3000

Instrument Cleaning

What Is an Ultrasonic Cleaner? How It Works, Sizes, Solutions & Testing

An ultrasonic cleaner is a tank of cleaning solution that uses high-frequency sound waves to remove blood, tissue and debris from instruments before they are sterilized. It cleans; it does not sterilize or disinfect. In dental, medical and lab settings it is the step between using an instrument and putting it in the autoclave.

The CDC recommends automated cleaning equipment, such as an ultrasonic cleaner or washer-disinfector, to improve cleaning and reduce staff exposure to blood, and notes that automated methods can be more efficient and safer than hand scrubbing. This guide covers how ultrasonic cleaners work, how to pick the right size, which solutions to use, how to test one, and how to care for it.

How does an ultrasonic cleaner work?

Transducers under the tank send high-frequency sound waves through the solution. The waves create microscopic bubbles that form and collapse rapidly, a process called cavitation. The collapsing bubbles loosen debris from instrument surfaces, including joints, hinges and crevices a brush can't reach. That's why public health guidance calls for ultrasonic cleaning of instruments with joints, crevices or lumens.

The cleaning solution does the chemical work: it breaks down blood, proteins and other soils so the cavitation can lift them away.

Where does an ultrasonic cleaner fit in instrument reprocessing?

Cleaning is the first step. Leftover debris can shield microorganisms and cause sterilization to fail, so instruments must be clean before they go into the sterilizer.

  1. Transport used instruments to the processing area in an appropriate container that prevents injuries.
  2. Clean in the ultrasonic cleaner, with the lid closed while it runs.
  3. Rinse off the cleaning solution, minimizing splashing.
  4. Dry and inspect each instrument for remaining debris. Re-clean anything that isn't clean.
  5. Package in sterilization pouches or wrap.
  6. Sterilize in the autoclave and monitor sterilization.

Important: the CDC says to handle instruments as contaminated until they are sterilized. Wear utility gloves, protective eyewear and a long-sleeved gown or jacket when loading, unloading, rinsing and packaging instruments.

How to choose an ultrasonic cleaner

Start with the size of your instruments and cassettes, then look at volume and installation.

  • Tank size: your longest instrument and largest cassette must lie flat in the basket, fully covered by solution. Check the manufacturer's tank dimensions, not just the gallon rating.
  • Workload: busy practices often size up so a full set of cassettes fits in one cycle.
  • Tabletop or recessed: recessed units drop into the countertop of a sterilization center to save space.
  • Drain: a drain valve lets you empty the tank without lifting and tipping it.
  • Timer and degas: a digital timer and a degas function make cycles repeatable.
  • Lid and basket: the lid should stay on while the unit runs to limit aerosols. Instruments should sit in a basket or cassette, not loose on the tank bottom.

Shop by brand

DuraSonic ultrasonic cleaners

DuraSonic

Made in the USA, in tank sizes from 0.5 to 6.6 gallons.

Shop DuraSonic →
SciCan StatClean ultrasonic cleaner

SciCan StatClean

1.6-gallon (6.0 L) tank, self-docking lid, hands-free drying basket and a 1–30 minute timer.

Shop StatClean →
Crest Powersonic ultrasonic cleaners

Crest Powersonic

Tabletop ultrasonic cleaners in multiple tank sizes.

Shop Crest →

Shop ultrasonic cleaners by size

Tank capacity Liters Shop
0.5 gallon 2.5 L 0.5 gallon cleaners
0.75 gallon 3 L 0.75 gallon cleaners
1 gallon 4 L 1 gallon cleaners
1.5 gallon 6 L 1.5 gallon cleaners, including the SciCan StatClean
2.1 gallon 8 L 2.1 gallon cleaners
2.5 gallon 9.5 L 2.5 gallon cleaners
3.25 gallon 12.3 L 3.25 gallon cleaners
4 gallon 15 L 4 gallon cleaners
5.25 gallon 20 L 5.25 gallon cleaners
6.6 gallon 25 L 6.6 gallon cleaners

Also see Tuttnauer Clean & Simple, BrandMax Tri-Clean recessed, or browse all ultrasonic cleaners.

Choosing the right cleaning solution

Use the type and amount of solution your cleaner's manufacturer specifies, such as an enzymatic cleaner, and mix it exactly as its label directs.

  • Check both instructions: the solution's label and your cleaner's manual. The CDC says to always follow the manufacturer's instructions for use. Some manuals also restrict water type; SciCan's StatClean manual, for example, says not to use osmosis or highly purified water.
  • Degas after every solution change. Fresh solution holds dissolved air that weakens cavitation. Public health guidance says to degas whenever the solution is changed or topped up.

How to use an ultrasonic cleaner, step by step

  1. Fill the tank between the minimum and maximum lines, then add the cleaning solution at the labeled dilution.
  2. Degas the fresh solution.
  3. Load instruments or cassettes in the basket, opened and fully submerged. Don't overload the basket.
  4. Close the lid and run the cycle time recommended by the cleaner and solution manufacturers.
  5. Remove, rinse and dry the instruments, wearing utility gloves.
  6. Inspect for remaining debris, then package and sterilize.

Never run the unit without solution in the tank. SciCan, for example, warns that running the StatClean dry for more than about 15 seconds can seriously damage it.

How to test an ultrasonic cleaner

An ultrasonic cleaner can sound normal and still not be cleaning. Test it regularly with a foil test or a commercial ultrasonic monitor, and log the results.

  • How often: public health guidance calls for efficacy testing at least weekly, or as the manufacturer recommends.
  • Foil test or test kit: an aluminum foil test or a commercial test kit shows whether cavitation is reaching the whole tank. Some manufacturers use their own version of the foil test, so follow your unit's instructions.
  • Commercial monitors: ultrasonic cleaner monitors give a quick visual result. Use the method in your cleaner's instructions where it differs.
  • If a test fails, repeat it with fresh solution. If it fails again, take the unit out of service and have it checked.

Caring for your ultrasonic cleaner

  • Change the solution at least daily, and sooner whenever it looks cloudy or soiled.
  • Clean the tank daily. At the end of each day, drain it, then clean and disinfect it as the manufacturer directs.
  • Use only the cleaning method in your unit's manual for the tank and housing. SciCan's StatClean manual, for example, warns against abrasive or strongly corrosive solutions.
  • Document maintenance the way the manufacturer's instructions call for.
  • Keep test records with your other instrument processing logs.

Shop ultrasonic cleaning at Duraline

Compare ultrasonic cleaners, cleaning solutions, baskets and monitors for professional instrument reprocessing, with financing available on equipment. Need help choosing a size? Call (877) 561-0500.

Shop Ultrasonic Cleaners →

Frequently asked questions

Does an ultrasonic cleaner sterilize instruments?

No. It only cleans. Instruments still have to be rinsed, dried, packaged and sterilized in an autoclave.

Is an ultrasonic cleaner better than hand scrubbing?

The CDC recommends automated cleaning equipment to improve cleaning and reduce staff exposure to blood, and says automated methods can be more efficient and safer than manual cleaning.

What size ultrasonic cleaner do I need?

One whose tank fits your longest instrument and largest cassette fully submerged. Compare the manufacturer's tank dimensions, not only the gallon rating, and consider how many cassettes you process per cycle.

What solution should I use?

The type and amount your ultrasonic cleaner's manufacturer specifies, mixed as the solution's label directs.

How often should I change the cleaning solution?

At least daily, and sooner whenever it looks cloudy or soiled.

What does degassing do?

It removes dissolved air from fresh solution so cavitation works at full strength. Degas every time you change or top up the solution.

How often should I test my ultrasonic cleaner?

At least weekly, or as often as your manufacturer recommends.

Should the lid be on while it runs?

Yes. Keep the lid on during use to limit aerosols and cross-contamination.

How should I clean the tank?

Drain it at the end of each day and clean and disinfect it using the method in your unit's manual.