Amsco Century Sterilizers are designed for ease of use with state of the art technology. The V148 model (Amsco Century Medium size) is designed for scientific applications with fast, efficient sterilization of heat and moisture-stable materials. It has cycles for pre-vacuum, gravity, liquid, leak test and daily air removal (Bowie-Dick) test cycles. With a large 26" x 37.5" x 48" chamber, this unit can come with a single door or double door with hinge or sliding options.
Features:
- Florescent display for easy reading
- Ink on paper impact printer
- 26″ x 37.5″ x 48″ chamber
- Pre-vacuum Steam Sterilizer
- Fast & efficient
- Emergency manual exhaust valve
- Non-lubricated, steam-activated door seal
- Help screens for programming/troubleshooting alarm conditions
- Communication interface compatibility with most PC devices
- Electronic water conservation valve
- Stainless steel side panels available for freestanding units.
Additional Options: Cabinet Package, Loading Cart, Integral Steam Generator, Water Filtration Options
Utility Requirements: Drain (D): 2" ODT drain terminal (floor drain
capacity must handle peak water
consumption)
Electric Control (EC) Domestic: 120 V, 1-phase, 60 Hz, 2 A
Electric Control (EC) International: 230 V, 50 Hz, 1-phase, 1.5 A
Electric Control, Vacuum Pump (VP)
Domestic: 480 V, 60 Hz, 3-phase, 3 A per
phase, or
208/240 V, 60 Hz, 3-phase, 6 A per
phase
Electric Control, Vacuum Pump (VP)
International: 400 V, 50 Hz, 3-phase, 4 A per
phase
Steam (S)
- Pressure – 50 to 80 psig (3.45 to
5.52 bar, dynamic), condensate free,
and 97% to 100% vapor quality
- Consumption – 185 lbs/hr (84 kg/
hour)
- Peak Flow – 335 lbs/hr (152 kg/hour)
- Size – 1" NPT
Sterilizer Feedwater (FW):
- Pressure – 20 to 50 psig (1.38 to
3.45 bar), dynamic
- Consumption – 130 gal/hr (495 L/
hour)
- Peak Flow – 15 gal/min (57 L/min)
- Size – 1" NPT
- Chamber Size: 26" x 37.5" x 48"
- Drain: 2" ODT drain terminal (floor drain capacity must handle peak water consumption)
- Electric:
- USA: Control: 120 volt, 1-phase, 60 Hz, 2 A. / Vacuum Pump: 208/240 volt, 3-phase, 60 Hz, 6 A per phase, or 480 volt, 3-phase, 60 Hz, 3 A per phase.
- UK: Control: 230 volt, 1-phase, 50 Hz, 1.5 A. / Vacuum Pump: 400 volt, 3-phase, 50 Hz, 6 A per phase.
- International: Control: 230 volt, 1-phase, 50 Hz, 1.5 A / (Outside UK) Vacuum Pump: 400 volt, 3-phase, 50 Hz, 4 A per phase.
- Steam:
- Pressure – 50 to 80 psig (3.45 to 5.52 bar, dynamic), condensate free, and 97 percent to 100 percent vapor quality
- Consumption – 185 lb/hr (84kg/hour)
- Peak Flow – 335 lb/hr (152 kg/hour)
- Size – 1" NPT
- Water:
- Pressure – 20 to 50 psig (1.38 to 3.45 bar), dynamic
- Consumption – 130 gal/hr (495 L/hr)
- Peak Flow – 15 gal/min (57 L/min)
- Size – 1" NPT
- Operating Weight: 4200 lb (1900 kg)
Steris Amsco V148 Steam Sterilizer Brochure Steris Chromalox Powerpack Boilers - Electric Steam Type CH02-CH14 Amsco Century Bio-Seal Devices - for double door Century small and medium steam sterilizers to separate work areas for biohazard or bioburden control (mostly in research laboratories and industrial facilities).
Standards:
• EMC Directive: 89/336/EEC, 92/31/ EEC, 93/68/EEC. • Low Voltage Directive: 72/23/EEC, 93/68/EEC. • Machine Directive: 89/392/EEC, 91/368/EEC, 93/44/EEC, 93/68/EEC. • Pressure Equipment Directive: 97/23/EC. • Underwriters Laboratory (UL) Standard 3101 as certified by ETL Testing Laboratories, Inc. • Canadian Standards Association (CSA) Standard C22.2, No. 1010. • ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form U-1 is furnished. Shell and door are constructed to withstand working pressure of 45 psig