Unique Point Class 5 Chemical Integrator for Steam sterilization, 200 Strips/pack
Couldn't load pickup availability
Product code: C5-INTG-SUP
OEM: IT26-1YS
Features
-
ISO 11140-1:2014 Type 5 Compliance:
- Conforms to Type 5 Indicators standards as per ISO 11140-1:2014, ensuring reliability in monitoring Steam sterilization processes.
-
Effective Sterilization Control:
- Developed for Steam sterilization processes between 121 ºC and 135 ºC, providing control over temperature, time, and steam quality.
-
Innovative Purple-to-Green Ink:
- Purple ink transitions to green when the theoretical spores population reaches its kill time, indicating successful integration conditions calibrated with a 106 Geobacillus stearothermophilus spore population.
-
High Sensitivity to Steam Variables:
- Manufactured with inks highly sensitive to Steam sterilization process variables, ensuring accurate and reliable results.
-
Versatile Size Options:
- Available in three variants: IT26S (72 mm x 23 mm paper strip), IT26SBL (two-sided laminated paper strip), and IT26SAD (self-adhesive paper strip).
-
Toxic Heavy Metals Free:
- 100% free of toxic heavy metals, prioritizing safety in sterilization processes.
-
Strategic Placement for Thorough Sterilization:
- Instructions recommend placement within each package, bag, or tray. In rigid containers, position in every corner or at least two diagonally opposite corners to ensure steam reaches inaccessible areas.
- Storage and Shelf Life:
- Store in dry places, away from light, at temperatures between 10-30 ºC, 30-80% relative humidity.
- Indicators have a shelf life of 5 years from the date of manufacture under recommended conditions.
- Do not use indicators after their expiration date.
- After use, the indicator remains stable for 6 months if stored correctly.
Instructions for Use:
- Place indicators within each package, bag, or tray.
- In rigid containers, place indicators strategically.
- After completing the process, analyze results. Integration condition is met if the ink changes to the reference color, indicating a successful sterilization process.
| Process | Steam |
|---|---|
| Packaging | 200 Strips/pack |
| Initial Color | Purple |
| Final Color | Green |
| Indicator Type | Type 5 Chemical Indicator |
| Regulations | ISO 11140-1 |
Instructions for Use:
- Place indicators within each package, bag, or tray.
- In rigid containers, place indicators strategically.
- After completing the process, analyze results. Integration condition is met if the ink changes to the reference color, indicating a successful sterilization process.