Bionova Super Rapid Self-Contained Biological Indicator for Steam sterilization processes with completed results in 1 hour, 50/box
FDA, Rapid Fluorescence
Product code: BI-BS-1
OEM: BT222
Features
Sold separately:
Bionova Incubator
Bionova MiniBio Auto-Reader Incubator
FDA, Rapid Fluorescence
Product code: BI-BS-1
OEM: BT222
Features
- 1-Hour Readout: 1 hour readout time for steam sterilization process.
- Spores: Geobacillus Stearothermophilus (ATCC® 7953) spores with a concentration of 10^6 spores per vial.
- Optional Visual pH Confirmation: users have the option to perform a visual pH color change confirmation after 48 hours, enhancing flexibility and offering an additional layer of confirmation.
- Culture Medium Color Indication: the culture medium undergoes a color change, initially to green and then yellow, if the sterilization process is unsuccessful. Conversely, a successful process maintains the medium in a purple hue, enabling clear differentiation.
- Dual D-Value Information: the product provides D-values for two temperatures (121 ºC and informed values at 132ºC and 135ºC), ensuring comprehensive data for different sterilization conditions.
- Fluorescence Detection Capability: the inclusion of fluorescence detection allows for an alternative and rapid assessment of sterilization effectiveness, adding an extra layer of reliability.
- Humidity Consideration for 7-Day Readout: when opting for the 7-day readout, users are advised to create a humidified environment to prevent the medium from drying out, ensuring accurate and reliable results.
- Shelf Life: 2 year expiration date from the date of manufacture under recommended storage conditions.
- User-Friendly Instructions: the product comes with clear and concise instructions for use, including labeling requirements, packaging guidelines, and post-sterilization procedures.
Sold separately:
Bionova Incubator
Bionova MiniBio Auto-Reader Incubator
Technical Specifications
| Process | Steam |
|---|---|
| Packaging | 50 Biological Indicators per box |
| Microorganism | Geobacillus stearothermophilus (ATCC® 7953) |
| Population | 10^6 spores per vial |
| Readout | 1 hour fluorescence readout at (60 ± 2) °C |
| Optional Visual Confirmation | 48 hours (pH color change) |
| Shelf Life | 2 years from the date of manufacture under recommended storage conditions |
| OEM | BT222 |
Storage Note: Store in a dark place within the temperature range of 10-30 °C and 30-80% relative humidity. Avoid freezing and keep biological indicators away from sterilizing agents or other chemical products.
Instructions for Use
- Labeling SCBI: Identify SCBI with sterilizer number, load number, and processing date.
- Packaging and Placement: Pack SCBI with materials for sterilization, placing it in less exposed areas.
- Sterilization Process: Carry out standard sterilization procedures.
- Post-Sterilization Steps: After sterilization, open the door, wait five minutes, and carefully remove SCBI. Caution: Wear safety glasses and gloves. Warning: Avoid excessive handling to prevent glass ampoule bursting. Let SCBI cool to room temperature.
- Process Indicator Check: Inspect the Process Indicator on SCBI label. Brown color indicates exposure to steam. Note: Color change does not confirm sterility; further checks needed if no change.
- Ampoule Crushing and Incubation: Seal the tube's lid and crush the ampoule using an ampoule crusher. Shake the tube to soak the spore carrier completely. Place SCBI in the incubator. Important: Use a non-sterilized SCBI as a positive control daily. Ensure both controls are from the same batch.
- Incubation and Readout: Incubate processed and positive control indicators for a maximum of 1 hour at (60 ± 2) °C. Note: Incubation should not exceed a 7-day period after sterilization. Failure in sterilization is indicated by fluorescence detection. Negative result if no fluorescence at 1 hour. Positive control must show positive fluorescence readout. Record positive results and discard SCBI promptly as instructed.
Precautions
- WARNING: Do not use SCBI to control sterilization processes other than those specified.
- WARNING: Place one or more SCBI in hard-to-reach areas to ensure complete chamber sterilization. Evaluate load configurations to identify and place SCBI in all such locations.
- WARNING: Do not reuse SCBI.
- WARNING: Do not reuse the sterilizer until the SCBI test result is negative.
FAQ
Scroll down for frequently asked questions
Scroll down for frequently asked questions
Explore Our FAQs for Helpful Answers
Find answers to common questions about this product, including features, specifications, and operation. Need more help? Our team is here to assist.
How long until I get results?
How long until I get results?
Up to 1 hour with fluorescence readout at (60 ± 2) °C in a Bionova auto-reader incubator. Need faster results? See the Ultra Rapid BI-BS-20 (20-minute results).
Which incubator do I need?
Which incubator do I need?
A Bionova auto-reader incubator, such as the Bionova MiniBio Auto-Reader Incubator.
What does the brown label mean?
What does the brown label mean?
The Process Indicator on the label turns brown when exposed to steam. This confirms exposure only, not sterility. If it has not changed, check the sterilization process.
Do I need a positive control?
Do I need a positive control?
Yes. Use a non-sterilized indicator from the same batch as a positive control daily. It must show a positive fluorescence readout.
Can I confirm results visually?
Can I confirm results visually?
Yes, optionally after 48 hours. The culture medium turns green and then yellow if sterilization failed, and stays purple if it was successful.
How should I store them, and how long do they last?
How should I store them, and how long do they last?
In a dark place at 10-30 °C and 30-80% relative humidity, away from sterilizing agents and chemicals. Do not freeze. Shelf life is 2 years from the date of manufacture.
How often should I spore test my sterilizer?
How often should I spore test my sterilizer?
The CDC recommends biological monitoring (spore testing) of each sterilizer at least weekly, and with every load that contains an implantable device. Your state or local regulations may require more frequent testing.
What should I do if a spore test fails (positive result)?
What should I do if a spore test fails (positive result)?
Take the sterilizer out of service and repeat the spore test right away using the same cycle. If the mechanical and chemical indicators show the sterilizer is working properly, a single positive result often points to operator error, such as loading or packaging. If the repeat test is also positive, do not use the sterilizer until it has been inspected or repaired and passes spore tests in three consecutive cycles.
Isn't the color change on the pouch or tape enough?
Isn't the color change on the pouch or tape enough?
No. Chemical indicators show that the sterilizing agent reached the item, but they do not prove sterilization was achieved. Only a biological indicator (spore test) confirms that highly resistant spores were killed.
Is a rapid-readout result as reliable as a 24- or 48-hour test?
Is a rapid-readout result as reliable as a 24- or 48-hour test?
Yes. Independent studies published in Infection Control & Hospital Epidemiology found rapid-readout biological indicators to be as sensitive as conventional 24- and 48-hour indicators, and the fluorescence results correlate with the 7-day visual readout.