Self-Contained Spore Ampoules for Monitoring Liquid Steam Sterilization, with Log 6 spore population, Size: 47mm X 10.6mm x 1ml (volume), 50/box
Couldn't load pickup availability
Important: This item must be shipped UPS OVERNIGHT on ice packs
Product code: BI-SA-L6
Spore ampule biological indicators can be used in Steam Autoclaves at 121°C to 137°C for monitoring liquid steam processes and dry loads. Must be used with Bionova IC10/20 incubator. For laboratory or industrial use only.
Note: Spore ampoules must be kept refrigerated and ship overnight. These are specially made for monitoring liquid sterilization in steam autoclaves.
Features- Log 6 spore population
- Size: 47mm x 10.6mm
- Volume: 1ml
- 50 per box
- Sterilization Type: Steam
- Loads: Liquid and Dry loads
Bionova IC10/20 Incubator
Technical Specifications
| Brand | True Indicating |
|---|---|
| Process | Steam |
| Loads | Liquid and Dry loads |
| Microorganism | Geobacillus stearothermophilus Cell Line 7953 |
| Spore Population | Log 6 |
| Size | 47mm x 10.6mm |
| Volume | 1ml |
| Temperature Range | 121°C to 137°C |
| Incubation | 60°C to 65°C for a minimum of 48 hours |
| Shelf Life | 24 months from the date of manufacture |
| Quantity | 50 per box |
| Incubator | Bionova IC10/20 |
Storage Note: Spore ampoules must be kept refrigerated and ship overnight. This item must be shipped UPS OVERNIGHT on ice packs.
Instructions for Use
- Place Spore Ampules inside representative materials (containers of liquid) or within the autoclave chamber directly. Package or wrap product as usual, if applicable. Locate product or Spore Ampules in most difficult location to sterilize, as outlined in your specific sterilization validation protocol or according to standard operating procedure.
- Run the cycle.
- After sterilization or exposure, remove Spore Ampules or product from sterilizer. Allow product or Spore Ampules to cool to the touch. No activation is required.
- Incubation: Place the processed Spore Ampules, one Negative Control Ampule and one Positive Control (unprocessed Spore Ampule) in a vertical position in an incubator at 60°C to 65°C for a minimum of 48 hours.
- Monitoring: Examine the Spore Ampules daily during incubation. Record observations. All positive Spore Ampules should be disposed of immediately. Do not continue to incubate a positive Ampule. Continued growth may result in metabolism of amino acids in the absence of sugars, causing the pH to rise and result in color reversion that is visibly darker than a sterile unit. These should be considered as positive for growth (turbidity will be present).
- Interpretation – Negative Control: The Negative Control Ampule should not exhibit a color change to Yellow and/or demonstrate turbidity. Utilize the Negative Control as a color comparison for the exposed Spore Ampules, where applicable.
- Interpretation – Positive Control Spore Ampule: The Positive Control Ampule should exhibit a color change to Yellow and/or demonstrate turbidity. Utilize the Positive Control as a color comparison for the exposed Ampules, where applicable. If the positive control does not demonstrate a Yellow color and/or turbidity, the results for the test Ampules should not be considered valid. Verify incubation conditions were met.
- Interpretation – Test Spore Ampules: A passing sterilization cycle is indicated by a test Spore Ampule which retains its original Purple color and is free from turbidity. A failed sterilization cycle is indicated by turbidity and/or a color change to Yellow.
- For unexpected positives, it is recommended that a Gram stain be performed. Gram positive rods are indicative for the indicator organism.
ISO 11138-1 Sterilization of health care products - Biological indicators - Part 1: General requirements
ISO 11138-3 Sterilization of healthcare products - Biological indicators - Part 3: Biological indicators for moist heat sterilization processes.
USP <55> Biological Indicators - Resistance Performance Tests
True Indicating has a validated method for Total Viable Spore Count. Please inquire for the Technical Bulletin which outlines the recommended methodology.
Scroll down for frequently asked questions
| Image | Product Code | Description | Spore Population | Size | Volume | Quantity | Price/box | View |
![]() | BI-MSA-ATM05 | Mini Spore Ampule | Log 5 | 40mm x 6.5mm diam | 0.7ml | 50/box | $220 | Add To Cart |
![]() | BI-MSA-ATM06 | Mini Spore Ampule | Log 6 | 40mm x 6.5mm diam | 0.7ml | 50/box | $220 | Add To Cart |
![]() | BI-SA-L4 | Spore Ampule | Log 4 | 47mm x 10.6mm | 1ml | 50/box | $220 | Add To Cart |
![]() | BI-SA-L5 | Spore Ampule | Log 5 | 47mm x 10.6mm | 1ml | 50/box | $320 | Add To Cart |
![]() | BI-SA-L6 | Spore Ampule | Log 6 | 47mm x 10.6mm | 1ml | 50/box | $339 | Add To Cart |