AN EO Gas 4PLUS Sterilizer
Andersen EOGas 4PLUS Ethylene Oxide Sterilizer
Product code: NANEOG4Plus
The EOGas 4PLUS is designed for terminal sterilization of difficult-to-sterilize medical devices, including duodenoscopes and colonoscopes. Powered by EO-Flexible Chamber Technology (EO-FCT), it uses a 17.6 g EO microdose—up to 90% less EO than traditional rigid-chamber systems.
The system supports active aeration in the same chamber, controlled-temperature air circulation, automatic door locking, and a touchscreen interface. When paired with the optional Andersen ETO Cartridge Abator, it provides near-zero-emissions performance for processing compatible heat- and moisture-sensitive instruments.
- FDA-cleared for terminal sterilization of complex long-lumen devices
- EO-FCT using a 17.6 g microdose of EO per cycle
- Near-zero emissions when paired with the Andersen ETO Cartridge Abator
- Approximately 3.5-hour cycle achieving a 10⁻⁶ sterility assurance level
- Active aeration in the same chamber with no device transfer required
- Touchscreen interface with real-time cycle guidance
- Automatic door locking during sterilization cycles
- Process Challenge Device integrated into the purge probe for consistent biological indicator placement
- Compared with steam: Supports compatible heat- and moisture-sensitive devices that may be damaged by steam.
- Compared with hydrogen peroxide (H₂O₂): Supports compatible complex and long-lumen devices.
- Compared with traditional EO systems: Uses a lower EO dose with a smaller equipment footprint and reduced emissions.
| Category | EOGas 4 | EOGas 4PLUS |
|---|---|---|
| Positioning | Proven EO sterilizer for long lumens and heat-sensitive devices | Next-generation EO platform for complex endoscopes and high-risk devices |
| Primary Use | General long-lumen and delicate-device sterilization | Terminal sterilization of duodenoscopes, colonoscopes, and complex lumens |
| FDA Clearance | Terminal sterilization of long-lumen devices | Terminal sterilization of duodenoscopes and colonoscopes |
| Chamber Technology | Traditional chamber | EO-Flexible Chamber Technology (EO-FCT) |
| EO Dose per Cycle | Low-dose EO | 17.6 g microdose EO |
| EO Efficiency | Reduced EO usage | Up to 90% less EO than rigid-chamber EO systems |
| Lumen Compatibility | Long-lumen capable | Validated for working lumens longer than 1,100 mm |
| Sterility Assurance Level | SAL 10⁻⁶ | SAL 10⁻⁶ |
| Cycle and Workflow | Standard EO cycle | Approximately 3.5-hour cycle with integrated aeration |
| Sterilization and Aeration | Aeration supported | Sterilization and aeration in the same chamber |
| Endoscope Optimization | Not endoscope-specific | Designed for flexible endoscopes, duodenoscopes, and colonoscopes |
| Environmental Impact | Low EO emissions | Near-zero emissions with the Andersen ETO Cartridge Abator |
| Process Monitoring | Standard monitoring | Process Challenge Device integrated into the purge probe |
| User Interface | Standard controls | Touchscreen with real-time cycle guidance |
| Safety Features | Standard door safety | Automatic door locking during the cycle |
| Temperature Control | Controlled chamber environment | Warm-air circulation for stable sterilization and aeration |
| Facility Fit | ASCs, veterinary facilities, SPD departments, and outpatient facilities | Hospitals, endoscopy centers, and advanced SPD departments |
| Cost Positioning | Cost-effective EO solution | Premium, compliance-focused system |
| Key Advantage | Proven reliability and lower acquisition cost | Lower EO usage and expanded device compatibility |
The EOGas 4PLUS is also cleared for sterilization of metal, plastic, and fabric loads.
Sterility Assurance Level: SAL 10⁻⁶
- Chamber Size: 25.25" L × 18" W × 14" H
- Exterior Dimensions: 29.5" L × 22" W × 28" H
- Weight: 161 lb (73 kg)
- Electrical: 115 V (installation supports a 120 V or 240 V outlet)
- EO Dose: 17.6 grams per cycle
| Exposure | Device Specifications | Maximum Load—Duodenoscopes | Maximum Load—Colonoscopes |
|---|---|---|---|
| 6-Hour Cycle |
Working lumen length greater than 1,100 mm
|
Two (2) duodenoscopes*
*One duodenoscope may be paired with one colonoscope.
|
Two (2) colonoscopes*
*One colonoscope may be paired with one duodenoscope.
|
| 3-Hour Cycle |
Working lumen length of 1,100 mm or less
|
One (1) endoscope
|
Four (4) endoscopes
|
EO Gas 4 FDA Clearance
Compare EO Gas models
A. EOGas 4Plus is an advanced gas sterilization system designed for efficiency and reliability. It utilizes a microdose of ethylene oxide to ensure sterilization of medical instruments. This system is ideal for facilities seeking a reliable solution to sterilization needs.
Q. How does EOGas 4Plus work?
A. The EOGas 4 PLUS operates by introducing ethylene oxide gas into a impermeable flexible chamber containing the devices to be sterilized. The gas penetrates packaging and surfaces, effectively killing bacteria and viruses. After a designated exposure time, the gas is evacuated, leaving devices terminally sterile with a shelf life of up to six months.
Q. Is it safe to use?
A. The chamber wash pump flow rate is100 L/mn
Q. What devices is EOGas 4PLUS intended for?
A. Heat- and moisture-sensitive devices, including duodenoscopes, colonoscopes, and long-lumen instruments.
Q. Does it require a separate aeration cabinet?
A. No. Sterilization and active aeration occur in the same chamber.
WARRANTY: 12 months parts only