Bionova MiniBio Auto-Reader Starter Kit with Ultra Rapid Biological Indicators, 20 Minute Results
Couldn't load pickup availability
MINIBIO Auto-Reader Incubator for Biological Indicators with 3 positions and built-in printer, with Ultra-Rapid Biological Indicators, 20 Minute results for Steam, Starter KIt
Product code: BI-BMBAR20-SK
Included with Kit:
For the incubation and automatic readout of Rapid, Super Rapid and Ultra Rapid Self-contained biological indicator (SCBIs) appropriate for sterilization and environmental disinfection process control applications.
MiniBio Auto-Reader Incubator Features:
-
Multi-Indicator Compatibility: Accommodates various Bionova / Tuttnauer Biological Indicators for comprehensive testing.
- 20 Minute & 1 hour process times can be used simultaneously
- Versatile Incubation: Three metal positions for simultaneous incubation of different indicators at customizable temperatures. (3 Incubation slots)
- Efficient Readout: Rapid, Super Rapid, and Ultra Rapid Fluorescence Readout options for swift results.
-
User-Friendly Design: Self-contained biological indicators and USB connectivity for easy result registration.
- Operating Specifications: Input 100/240V 0.2A (50/60Hz)/Output: 12V 4A
-
Dual Temperature System: Selectable between 37ºC and 60ºC for optimal indicator performance.
- Working Temperature: 37 ± 2°C, 60 ± 2°C
- Integrated Safety Features: Sound alarm, automatic cancellation of fluorescence reading, and event indication.
- Compact and Convenient: Dimensions of 180mm x 154mm x 105mm for space-saving efficiency.
- Upgradable Firmware: User-friendly firmware update capability for future enhancements.
Bionova Ultra Rapid Biological Indicators Features
-
Rapid Results:
- 20-minute readout time for quick sterilization process feedback.
-
High Spore Concentration:
- Contains Geobacillus Stearothermophilus (ATCC® 7953) spores at 10^6 spores per vial.
-
Optional Visual pH Confirmation:
- pH color change confirmation after 48 hours.
-
Cultural Medium Color Indication:
- Green to yellow color change for unsuccessful sterilization; purple for successful sterilization.
-
Dual D-Value Information:
- Provides D-values for 121 ºC, and informed values at 132ºC and 135ºC.
-
Fluorescence Detection Capability:
- Alternative rapid assessment of sterilization effectiveness.
-
Humidity Consideration for 7-Day Readout:
- Requires humidified environment to prevent medium drying out.
-
Shelf Life:
- 2-year expiration date under recommended storage conditions.
-
User-Friendly Instructions:
- Includes clear usage instructions, labeling, packaging guidelines, and post-sterilization procedures.
Purchase separately:
Rapid, Super Rapid or Ultra Rapid biological indicators
Autoclave LogBook for Steam Sterilizers
MINIBIO Auto-Reader Incubator
Ultra Rapid Self-Contained Biological Indicator
- 50 Biological Indicators per box
Storage Guidelines
- Store in a dark place within the temperature range of 10-30 °C and 30-80% relative humidity.
- Avoid freezing and keep biological indicators away from sterilizing agents or other chemical products.
Instructions for Use: Bionova Ultra Rapid Biological Indicators
Labeling SCBI:
- Identify SCBI with sterilizer number, load number, and processing date.
Packaging and Placement:
- Pack SCBI with materials for sterilization, placing it in less exposed areas.
Sterilization Process:
- Carry out standard sterilization procedures.
Post-Sterilization Steps:
- After sterilization, open the door, wait five minutes, and carefully remove SCBI.
- Caution: Wear safety glasses and gloves.
- Warning: Avoid excessive handling to prevent glass ampoule bursting. Let SCBI cool to room temperature.
Process Indicator Check:
- Inspect the Process Indicator on SCBI label. Brown color indicates exposure to steam.
- Note: Color change does not confirm sterility; further checks are needed if no change.
Ampoule Crushing and Incubation:
- Seal the tube's lid and crush the ampoule using an ampoule crusher.
- Shake the tube to soak the spore carrier completely.
- Place SCBI in the incubator.
- Important: Use a non-sterilized SCBI as a positive control daily. Ensure both controls are from the same batch.
Incubation and Readout:
- Incubate processed and positive control indicators for a maximum of 20 minutes at (60 ± 2) °C.
- Note: Incubation should not exceed a 7-day period after sterilization.
- Failure in sterilization is indicated by fluorescence detection.
- Negative result if no fluorescence at 20 minutes.
- Positive control must show positive fluorescence readout.
- Record positive results and discard SCBI promptly as instructed.
- WARNING: Do not use SCBI to control sterilization processes other than those specified.
- WARNING: Place one or more SCBI in hard-to-reach areas to ensure complete chamber sterilization. Evaluate load configurations to identify and place SCBI in all such locations.
- WARNING: Do not reuse SCBI.
- WARNING: Do not reuse the sterilizer until the SCBI test result is negative.
Q: How should I clean the external surfaces of the MiniBio Auto-reader?
A: Disconnect power and USB cables, wait for cooling. Use a damp microfiber cloth with mild detergent solution. Wipe, then rinse with water. Air dry for at least 1 hour before reconnecting cables.
Q: When should I perform the cleaning procedure?
A: Clean externally whenever spills occur or as considered appropriate. It's crucial after any spill on the device.
Q: Can I clean the internal parts of the Auto-reader?
A: No, do not clean internal parts. Contact your local distributor if further cleaning is needed or if unsure about cleaning agents.
Q: What if additional cleaning is required or I'm uncertain about cleaning agents?
A: Contact your local distributor for guidance.
Maintenance and Warranty
Q: Does the MiniBio Auto-reader require routine maintenance?
A: No, routine maintenance is not required. **
Q: What does the warranty cover, and for how long?
A: Terragene guarantees material and manufacturing quality. Warranty is 1 year from the date of first use, not extending beyond 5 years from manufacture.
Q: What is the procedure if a material or manufacturing fault is detected during the warranty period?
A: Terragene will repair or replace the product if material or manufacturing faults are detected within the warranty period.
Q: What is Terragene®'s liability in case of loss or damage resulting from equipment use?
A: Duraline Systems Inc. and Terragene are nnot liable for loss or damage due to unsuitable use, negligence, or user's full responsibility.
Technical Assistance
Q: How can I contact Duraline Systems for technical assistance?
A: Yes
Maintenance and Warranty Q: Does the MiniBio Auto-reader require routine maintenance? A: No, routine maintenance is not required.
** Q: What does the warranty cover, and for how long? A: Terragene guarantees material and manufacturing quality.
Warranty is 1 year from the date of first use, not extending beyond 5 years from manufacture.
Q: What is the procedure if a material or manufacturing fault is detected during the warranty period? A: Terragene will repair or replace the product if material or manufacturing faults are detected within the warranty period.